Lateral Flow Assay

Lateral Flow Assays, also known as Lateral Flow Immunochromatographic Assays, are simple devices that detect the presence (or absence) of a target analyte within a sample, without the need for specialised and costly equipment. Numerous applications have been developed for the specific qualitative, or semi-quantitative, detection of antigens and antibodies. Typically, these tests are used for medical diagnostics either for home testing, point of care testing, or laboratory use - a widely spread, and well known, application is the home pregnancy test. The technology is based on a series of capillary beds; such as pieces of porous paper, micro-structured polymer, or sintered polymer. Each of these elements has the capacity to transport fluid (e.g., urine), spontaneously.

See below for a one-step Lateral Flow Assay protocol, which allows for the rapid identification of various analytes. The protocol describes the preparation of the assay for the detection of the Infectious Bursal Disease Virus (IBDV) and Trichinella-specific antibodies.

A. Reagents and Equipment

Stock Solutions

  • Hydrogen tetrachloroaurate (1% (w/v)): Dissolve 0.1 g of HAuCl4 in 10 ml of triple-distilled water and store in a brown bottle.
  • Trisodium citrate (1% (w/v) ): Dissolve 0.5 g of Na3C6H5O7·2H2O in 50 ml of triple-distilled water.
  • NaCl (1 M): Dissolve 5.8 g of NaCl in 100 ml of triple distilled water.
  • K2CO3 (0.2 M): Dissolve 2.8 g of K2CO3 in 100 ml of triple-distilled water.
  • Sodium borate (20 mM): Dissolve 0.76 g of Na2B4O7·10H2O in 100 ml of triple-distilled water.
  • Phosphate buffered saline (PBS): Dissolve 8 g NaCl, 0.2 g KCl, 1.44 g Na2HPO4, 0.24 g KH2PO4 in 1,000 ml of triple-distilled water.

Materials and Equipment

  • Nitrocellulose Membrane
  • Glass Fibre Conjugate Pad Sheets (to make sample and conjugate pads).
  • Filter paper (to make absorbent pads).
  • Double-sided adhesive tape (to make adhesive layers on the backing plate).
  • Plastic films with different colours (to make strip covers).
  • Plastic boards (use for strip backing plate).
  • The XYZ 3000 platform with BioJet Quanti 3000 and AirJet Quanti 3000 dispensers (BioDot Inc., Irvine, CA) is used for membrane blotting and conjugate dispensing.
  • Forced Air Drying Oven (used to dry the blotted membrane and conjugate pad).
  • Laminator (to assemble the master card of test strips).
  • Guillotine Cutter (to cut the assembled master cards into strips).

B. Preparation and Procedure

Preparation of Colloidal Gold Suspension

  1. Add 1 ml of 1% (w/v) HAuCl4 solution in 100 mL of triple distilled water to a clean 500 ml Erlenmeyer flask on a stirring hot plate and bring the solution to a boil.
  2. To the boiling solution, quickly add 1 mL of 1% (w/v) of trisodium citrate under constant stirring.
  3. After the colour of the solution has changed from blue to dark red (within 3 min), continue to boil the solution for a further 2 min.
  4. When the solution cools to room temperature (RT), add triple-distilled water up to the original volume, stopper the flask to prevent evaporation, air circulation, and entry of contaminants and store the sample away from light.
  5. Scan the optical density of the red solution between 500 and 600 nm, and the absorption maximum should be at 525 nm, indicating that the gold particles have an average diameter of 40 nm.
  6. The colloidal gold suspension can be stored at RT in the dark for several months.

Conjugation of IgG with Colloidal Gold

for IBDV Antigen Detection for anti-Trichinella Antibody Detection
  1. Adjust the pH of the colloidal gold suspension to 9.0 with 0.2 M K2CO3.
  2. Dialyse the anti-IBDV mAb IgG against 100 volumes of 20 mM sodium borate for 24 h at 4°C with three changes of solution over this period.
  3. Add 2 mL of the IgG solution at the predetermined optimum concentration into 10 mL of the colloidal gold solution (pH 9.0), rapidly mix, and then incubate for 20–40 min at RT.
  4. Add 1/10 volume of 10% (w/v) BSA in 20 mM sodium borate. Mix rapidly upon addition and incubate for 10–15 min at RT.
  5. Centrifuge for 30 min at 15,000 × g at 4°C and carefully discard the supernatant.
  6. Re-suspend the pellet in 20 mM sodium borate containing 1% (w/v) BSA.
  7. Centrifuge again, discard supernatant, and re-suspend the pellet in 20 mM sodium borate containing 1% (w/v) BSA.
  8. Wash one more time and finally re-suspend the conjugate in 1 mL of 20 mM sodium borate containing 1% (w/v) BSA.
  1. Adjust the pH of the stock colloidal gold solution to 8.5 with 0.2 M K2CO3.
  2. Dialyse the ES antigens against 20 mM sodium borate at 4°C.
  3. Add 2 ml of the IgG solution at the predetermined optimum concentration into 10 ml of the colloidal gold solution (pH 9.0), rapidly mix, and then incubate for 20–40 min at RT.
  4. Add 1/10 volume of 10% (w/v) BSA in 20 mM sodium borate. Mix rapidly upon addition and incubate for 10–15 min at RT.
  5. Centrifuge for 30 min at 15,000 × g at 4°C and carefully discard the supernatant.
  6. Re-suspend the pellet in 20 mM sodium borate containing 1% (w/v) BSA.
  7. Centrifuge again, discard supernatant, and re-suspend the pellet in 20 mM sodium borate containing 1% (w/v) BSA.
  8. Wash one more time and finally re-suspend the conjugate in 1 ml of 20 mM sodium borate containing 1% (w/v) BSA.

Membrane Blotting

for IBDV Antigen Detection
for anti-Trichinella Antibody Detection
  1. Cut nitrocellulose membranes into 2.5 × 30 cm2 tapes.
  2. Dilute the other anti-IBDV mAb IgG and goat anti-mouse IgG to 1 mg/mL in PBS (pH 7.2) for the test and control lines, respectively.
  3. Put membrane tapes on the XYZ 3000 platform. Dispense the solutions onto the membranes at 1 μL/cm as test and control lines in parallel using the BioJet Quanti 3000. The test and control lines are located at the centre of the membrane with a space of 0.5 cm.
  4. Dry the blotted membranes at 42°C for 30 min in the Forced Air Drying Oven.
  5. Seal the membranes in a plastic bag in the presence of desiccants and store at 2–8°C.
  1. Cut nitrocellulose membranes into 2.5 × 30 cm2 tapes.
  2. Dilute the SPA and goat anti-ES IgG solutions to 1 mg/mL in PBS (pH7.2).
  3. Filter the solutions using 0.22 μm filters.
  4. Put membrane tapes on the XYZ 3000 platform. Dispense the solutions onto the membranes at 1 μL/cm as test and control lines in parallel using the BioJet Quanti 3000. The test and control lines are located at the centre of the membrane with a space of 0.5 cm.
  5. Dry the blotted membranes at 42°C for 30 min in the Forced Air Drying Oven.
  6. Seal the membranes in a plastic bag in the presence of desiccants and store at 2–8°C.

Preparation of Conjugate Pads

  1. Cut fibreglass into 1.5× 30 cm2 strips.
  2. Add 1 mL of the conjugate in 2 mL of 20 mM sodium borate (pH8.0) containing 2% (w/v) BSA, 3% (w/v) sucrose, 0.6 M NaCl, 0.2% (v/v) Tween 20, and 0.1% (w/v) sodium azide.
  3. Put the fibreglass strips on the XYZ 3000 platform. Dispense the conjugate onto the fibreglass at 15 μl/cm using the AirJet Quanti 3000.
  4. Dry the conjugate pads at 50°C for 30 min using the Forced Air Drying Oven.
  5. Seal the strips with desiccants in a plastic bag and store at 2–8°C.

Preparation of Sample Pads

  1. Cut fibreglass into 1.5 × 30 cm2 strips.
  2. Soak the strips in PBS (pH 7.2) containing 0.1 M NaCl, 0.2% (v/v) Tween 20, and 0.1% (w/v) sodium azide.
  3. Dry the strips at 50°C for 30 min using the Forced Air Drying Oven.
  4. Seal the sample pads with desiccant in a plastic bag and store at RT.

Preparation of Absorbent Pads

  1. Cut filter paper in 2.5 × 30 cm2 strips.
  2. Seal the absorbent pads with desiccant in a plastic bag and store at RT.

Preparation of Adhesive Cards

  1. Put double-sided adhesive tapes on one side of the support cards.
  2. Cut the adhesive board into 7.5 × 30 cm2 strips to make adhesive cards.

Assembly of Master Card

  1. Use a laminator to or manually assemble the pre-cut materials into a lateral flow strip master card (Fig 1. A
  2. Stamp the blotted membrane in the middle of the adhesive backing card. Then sequentially affix the conjugate and sample pads next to the membrane with a 1–2 mm overlap at the sample end, and the absorbent pad to distal end of the membrane with 1–2 mm overlap (Fig 1. B).
  3. Cover the sample and conjugate pads at the sample end, and the absorbent pad at the distal end with white and blue cover films, respectively (Fig 1. C).

Fig 1. Components for the strip lamination: 1. white cover film with arrows (sample end), 2. sample pad, 3. conjugate pad, 4. blotted membrane, 5. absorbent pad, 6. blue cover film (distal end), 7. adhesive backing plate.

Cutting and Packaging

  1. Cut the assembled master card into 0.3 cm strips using a guillotine cutter.
  2. Seal the test strips with desiccants in a plastic package and store at 2–8°C.

Detection

for IBDV Antigen Detection
for anti-Trichinella Antibody Detection
  1. Simply break up the bursal tissue in a whirl pack with finger massage, and dilute in PBS or water.
  2. Dip the IBDV antigen test strip into the solution for 10–20 s at RT.
  3. Take the strip out and place horizontally for 1–2 min to observe the result.
  4. If both the test and control lines turn red, the sample is recorded as positive, indicating the presence of IBDV antigens in the sample. When the control line but not the test line is coloured, it is considered as negative. No coloured band appears on the membrane indicates an improper testing procedure or deterioration of the strip, and the test should be repeated using a new strip
  1. Dilute blood or serum samples of human, pigs, cats, or dogs in PBS or water at 1:100.
  2. Dip the ES-antibody test strip into the solution for 10–20 s at RT.
  3. Take the strip out and place horizontally for 1–2 min to observe the result.
  4. If both test and control lines appear in red, the sample is recorded as positive, indicating the presence of antibodies specific to ES antigens. Appearance of only the control line but not the test line is considered as negative. No coloured line appearing on the membrane indicates an improper testing procedure or deterioration of the strip, and the test should be repeated using a new strip.